Abstract

Coronary stent restenosis, which emerges in late periods after implantation, has not been completely abolished. Our aim was to investigate the restenosis rates of Ephesos coronary stents. In all, 96 patients (66 men) with 135 Ephesos coronary stents were included. Control angiograms were performed after 160 +/- 60 days. Quantitative coronary analysis was performed during the procedure and control angiogram. The stents were divided into 2 groups according to the presence or absence of restenosis. Groups were compared with clinical and angiographic variables. Restenosis was observed in 31 (23%) of 135 stents. Preprocedure percent diameter stenosis was higher (P = .02), whereas minimum lumen diameter ( P = .02), mean age (P < .001), and hypertension incidence ( P = .043) was less, and there was a trend toward smaller stent size ( P = .054) in the restenosis group. By multivariate analysis, age <50 years (P < .001) and stent size <3.0 mm (P = .016) were independent predictors of restenosis. Ephesos coronary stents seems to have acceptable restenosis rates.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.