Abstract

Although several studies have shown positive outcomes after the use of drug-coated balloons (DCB) for in-stent restenosis (ISR), data on randomised controlled trials versus latest-generation drug-eluting stents (DES) are limited. Therefore, in this randomised trial, we sought to evaluate whether a butyryl-tri-hexyl citrate (BTHC)-based paclitaxel DCB is non-inferior to a biodegradable polymer sirolimus-eluting stent (BP-SES) therapy in patients with ISR in either a bare metal stent (BMS) or DES. A total of 229 patients with ISR in BMS or DES from 13 German centres and one Latvian centre were 2:1 randomly allocated to DCB (n=157) or DES (n=72). The primary efficacy endpoint was defined as in-stent late lumen loss (LLL) at six months, and the primary safety endpoint was target lesion failure (TLF) at 12 months. LLL in the DCB arm was 0.03±0.40 mm compared to 0.20±0.70 mm in the DES arm (p=0.40). DCB proved to be non-inferior to DES (Δ = -0.17±0.52 mm, 97.5% CI -∞; -0.01]; p<0.0001). At 12 months, Kaplan-Meier TLF estimates were 16.7% in the DCB arm and 14.2% in the DES arm (p=0.65) and remained similar at 18 months (DCB versus DES: 17.4% versus 19.5%, p=0.88). In patients with DES or BMS ISR, treatment with a paclitaxel DCB showed similar LLL at six months and TLF rates up to 18 months compared to a second-generation sirolimus DES.