An aneurysm of a 14-mm Contegra bovine conduit 5 years after a total repair of tetralogy of Fallot was confirmed by echocardiography, angiography, and magnetic resonance tomography. The conduit was replaced. Histologic examination of the explanted conduit revealed an acellular homogenous material with occasional elastic fibers, fragile, diffuse and complex collagenization throughout the conduit and mild foreign body reaction. Pannus formed over the top of all commissures and on the conduit wall, with extensive mineralization. Close follow-up is seen as mandatory for early detection of the bovine vein conduit aneurysm, particularly in patients in whom small-sized conduits are implanted.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.