Abstract

The Pipeline Vantage Embolization Device is a fourth-generation flow diverter with an antithrombotic coating and a reduced profile compared to previous Pipeline versions. The objective of this study was to evaluate the procedural feasibility, safety, and efficacy of this device. The Pipe-VADER study was designed as a retrospective, observational study of consecutive patients treated with the Vantage at 3 neurovascular centers. Patient and aneurysm characteristics, procedural parameters, early complications, and extent of postinterventional contrast retention were analyzed on an intention-to-treat basis. Twenty-eight patients with 31 aneurysms (median size: 5.0mm, posterior circulation: 4 [12.9%], ruptured: 5 [16.1%]) were included. The technical success rate was 100%, with multiple stents used in 4/30 (13.3%) procedures. Of the 30 procedures, adjunctive coiling was performed in 3 (10.0%) and balloon angioplasty in 2 (6.7%). Median procedure time was 62minutes. Procedural ischemic stroke occured in 4 (13.3%) cases, whereof 2 were major strokes (6.6%). There were no hemorrhagic complications. Initial contrast retention was observed in 29/31 (93.5%) aneurysms. All 27 overstented side vesselswere patent at the end of the procedure. Short-term follow-up (median: 5 months) showed complete and favorable occlusion rates of 70% (14/20) and 80% (16/20), respectively. The new Pipeline Vantage appears to be safe and feasible for the treatment of intracranial aneurysms and warrants further evaluation.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call