Abstract

The purpose of this study is to evaluate anatomy based inverse planning as implemented in PLATO BPS 14.2 for planning of HDR prostate implants. Six patients were analysed. The dose distributions were optimized using geometric optimization followed by graphical optimization (GO), anatomy based inverse planning or standard inverse optimization (SIO), tuned inverse optimization (TIO) and tuned inverse optimization followed by graphical optimization (GOTIO). The mean target coverage was 93±4%, 53±11%, 74±8%, 90±3%, respectively, for GO, SIO, TIO and GOTIO. The conformal index COIN was 0.74±0.02, 0.43±0.15 and 0.77±0.07, respectively, for GO, SIO and GOTIO. Improved dose homogeneity was found when comparing GOTIO with GO.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.