Abstract

The purpose of this study was to evaluate the changes of noise and computer tomography (CT) number in each dental substance, by using the metal artefact reduction algorithm; we used dual CT for this study. For the study, we produced resin, titanium, gypsum, and wax that are widely used by dentists. In addition, we made nickel to increase the artefact. While making the study materials, we made sure that there is no difficulty when inserting the substances inside phantom. In order to study, we scanned before and after using the metal artefact reduction algorithm. We conducted an average analysis of CT number and noise, before and after using the metal artefact reduction algorithm. As a result, there was no difference in CT number and noise before and after using the metal artefact reduction algorithm. However, when it comes to the noise value in each substance, wax's noise value was the lowest whereas titanium's noise value was the highest, after applying the metal artefact reduction algorithm. In nickel, CT number and noise value from artefact area showed a decreased noise value when applying the metal artefact reduction algorithm. In conclusion, we assumed that we could increase the effectiveness of CT examination by applying dual energy's metal artefact reduction algorithm.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.