Abstract
Thyroid stimulating hormone (TSH) immunoassays are known for giving varying results based on the platform of testing and the generation of kit used. It is generally expected that the results should not vary to affect clinical diagnosis and management. We aimed to perform method comparison study between two TSH assays by the same manufacturer Siemens Healthineers. Results show that there is a large proportional error between the assays with a bias of -3.71mIu/L indicating that TSH assay gives higher values for TSH for the same patient as measured against the TSH3-Ultra kit. This can affect interpretation of results leading to false increase in patients categorized under hypothyroidism and subclinical hypothyroidism. We strongly suggest, to prevent errors in clinical evaluation of a patient with thyroid dysfunction, validation of the performance of the assay and method comparison should be performed in-house.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.