Abstract

The analytical validation of multiple lots of the PRO-Trac II ELISA (DiaSorin) for the determination of tacrolimus in whole blood is described. The analytical parameters assessed included analytical sensitivity, dilution linearity, functional sensitivity, values in samples containing no tacrolimus, intra- and interassay precision, supplementation and recovery, metabolite cross-reactivity, interference studies, and method comparisons HPLC-tandem mass spectrometry (HPLC/MS/MS) and the IMx Tacrolimus II multiparticle enzyme immunoassay. Where appropriate, assessments were performed according to NCCLS guidelines. The mean analytical detection limit was <0.25 microg/L for all lots, whereas the functional sensitivity was 1.0 microg/L. Excellent linear correlation (r = 0.985) was observed for dilution linearity. The intraassay imprecision was <7%, and the total imprecision by ANOVA was <10%. Recovery was 109% +/- 11%. Metabolite cross-reactivity was consistent with previous reports for this antibody. No interference was observed for 35 tested drugs. Method comparison with HPLC/MS/MS showed no statistically significant differences. Samples exhibited stability through four freeze/thaw cycles and for 1 week at room temperature. These data demonstrate that the PRO-Trac II ELISA is a robust, accurate, and precise tool for the assessment and management of tacrolimus blood concentrations in transplant patients.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.