Abstract

The heavy/light chain (HLC) immunoassay quantifies the different heavy chain/light chain combinations of each immunoglobulin (Ig) class. This makes the HLC assay suited to quantify monoclonal immunoglobulins (M-protein) and for monitoring of patients with monoclonal gammopathies. This method is particularly advantageous for those samples in which electrophoretic quantification of the M-protein is not possible. In this study we tested the analytical performance of the HLC assay in 166 routine clinical samples and in 27 samples derived from the Dutch external quality assessment (EQA) for M-protein diagnostics (74 participating laboratories). Analytical accuracy was assessed by verification that the sum of the HLC-pairs equaled total Ig concentration. Sensitivity of the HLC assay was determined in a direct method comparison with immunofixation electrophoresis (IFE). Comparison of HLC data with routine Ig diagnostics in 27 EQA samples showed very good correlation for both the quantification of polyclonal and monoclonal IgG, IgA and IgM (Pearson correlations [r] were 0.94, 0.99 and 0.99, respectively; slopes were 0.94, 1.07 and 0.98, respectively). The overall concordance between IFE and the HLC ratio was high (93%) with a Cohen κ coefficient of 0.84. Discrepancies between both assays were mainly caused by the higher sensitivity of IFE to detect monoclonality. We conclude that the HLC assay is an accurate method to quantify M-proteins that can improve monitoring of M-proteins in the beta fraction that cannot be quantified using electrophoretic techniques.