Abstract

Preeclampsia is one of the leading hypertensive disorders of pregnancy. Angiogenic biomarkers such as anti-angiogenic factor soluble fms-like tyrosine kinase 1 (sFlt1) and pro-angiogenic factor placental growth factor (PlGF) are involved in the pathophysiology of preeclampsia. The aim of this study is to validate the analytical performance of sFlt1 and PlGF on the B·R·A·H·M·S KRYPTOR Compact Plus (ThermoFisher Scientific). We examined K2-EDTA plasma samples from 50 patients on B·R·A·H·M·S KRYPTOR Compact Plus, an automated immunoassay platform. QC materials were used to assess intra- and inter-precision of the assay. Lower limit of quantitation and interference studies were determined using pooled patient plasma. The sFlt1 and PlGF assays demonstrated an analytical measuring range of 90-69,000 pg/mL and 11-7000 pg/mL, respectively (r2 > 0.99). Lower limit of quantitation (20% CV) was interpolated to be 35 pg/mL for sFlt1 and 10 pg/mL for PlGF. Total precision for both assay displayed CVs of <10%. Interference studies showed that both assays were not significantly affected by hemolysis up to an H-index of 1100 for sFlt1 and 300 for PlGF; L- and I-index of 800 and 80 respectively for both assays. The Passing-Bablok regression analysis for sFlt1/PlGF yielded an equation of y = 1.05x + 0.02, and the Bland Altman analysis showed an average bias of 0.84. Plasma levels of sFlt1 and PlGF measured on the B·R·A·H·M·S KRYPTOR Compact Plus platform demonstrate excellent analytical performance and are acceptable as clinical grade assays.