Abstract

There are often discrepancies when using different methods to measure anti-Toxoplasma gondii IgG levels in patient samples. The diagnostic performance of a chemiluminescent immunoassay (CLIA) and an enzyme-linked fluorescent assay (ELFA) used as confirmatory tests for samples identified as positive or equivocal by an electrochemiluminescent immunoassay (ECLIA) were examined. Cut-off values were those stated by the manufacturer, and Western blot was used to confirm the results of all methods. All samples identified as positive by ECLIA (n=93) were confirmed as positive by Western blot, as were 14 of the 28 samples identified as equivocal. When these 121 samples were retested, the sensitivities of CLIA and ELFA were 64.4% and 73.8%, respectively. Both methods exhibited a specificity of 100%. This study confirms that the results obtained from the different immunoassays are not comparable, and neither CLIA nor ELFA should be used to confirm ECLIA results, which should instead be confirmed by methods such as Western blot or Sabin-Feldman dye test.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.