Abstract

Determining appropriate analytical thresholds (ATs) for forensic DNA analysis is critical to maximize allele detection. In this study, six methods to determine ATs for forensic DNA purposes were examined and compared. Four of the methods rely on analysis of the baseline noise of a number of negatives, while two utilize the relationship between relative fluorescence unit signal and DNA input in the polymerase chain reaction (PCR) derived from a dilution series ranging from 1 to 0.06 ng. Results showed that when a substantial mass of DNA (i.e., >1 ng) was amplified, the baseline noise increased, suggesting the application of an AT derived from negatives should only be applied to samples with low levels of DNA. Further, the number and intensity of these noise peaks increased with increasing injection times, indicating that to maximize the ability to detect alleles, ATs should be validated for each post-PCR procedure employed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.