Abstract

Abstract Two new liquid membrane electrodes which respond to creatininium and tetramethylammonium cations are described. The creatininium cation electrode exhibits rapid and near Nernstian response to creatininium cation activity, at pH 3, in the 10−3−10−1 mol/L range. The useful concentration range extends to 10−4 mol/L. The tetramethyl-ammonium cation electrode exhibits rapid and near Nernstian response to tetramethylammonium cation activity, at pH 2–11.5, in the 2x10−5− 10−1 mol/L range. Major interferences for the creatininium electrode are Na+, K+, NH+ 4 and creatine. The pKa of the creatininium cation was calculated. A method is described for the potentiometric precipitation titration of tetramethylammonium cation with sodium tetraphenylboron. Amounts of tetramethylammonium in the range 20–200 μol have been determined using Gran's plots, with an average error of about 0.6%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.