Abstract

Aims: The main aim of our study is to analyse Electron Beam Profiles like Flatness, Symmetry with 2D Array Detectors. The 2D Array can be used as an alternative device to measure the Electron beam profiles. Introduction: Flatness and symmetry of a radiation beam: The flatness of the beam is defined by the following formula: Flatness (%) = Dmax/Dmin × 100% (1) Where Dmax and Dmin are the maximum and minimum doses respectively within the area. Radiation field symmetry is defined as the maximum ratio of doses at two symmetric points relative to the central axis of the field. Symmetry (%) = [D(x, y)]/[D (-x,-y)] × 100% (2) Methods and Material: The beam symmetry is easily defined and is not very dependent on the depth of measurement. However, the flatness of the beam depends on the size and shape of the measurement phantom. However, 3% is sufficient measurement accuracy for a quick check. We measured the electron profiles of energies (4 MeV, 6 MeV, 8 MeV, 10 MeV, 12 MeV, 15 MeV and 18 MeV) from Elekta Synergy Linear accelerator for different electron applicators (6 × 6, 10 × 10, 14 × 14 and 20 × 20) at their depth of maximum dose (dm) respectively and Source-Surface-Distance (SSD) of 100 cm using PTW 729 2D array detectors. Results: We analyzed the variations in Flatness and Symmetry of electron energies obtained from 729 2D Array Detector and compared with the standard values obtained from the Radiation Field Analyzer (RFA) during Commissioning. Conclusions: We found that there is no significant variation in Flatness and symmetry obtained from the 729 2D Array detector as compared to the standard Flatness and symmetry obtained from the RFA .Thus we conclude that 729 2 D Array detector can be used for the routine measurement of electron beam profiles.

Highlights

  • IntroductionFlatness and symmetry of a radiation beam: The flatness of the beam is defined by the following formula: Flatness (%) = Dmax/Dmin × 100% (1)

  • Flatness and symmetry of a radiation beam: The flatness of the beam is defined by the following formula: Flatness (%) = Dmax/Dmin × 100% (1)Where Dmax and Dmin are the maximum and minimum doses respectively within the area

  • We analyzed the variations in Flatness and Symmetry of electron energies obtained from 729 2D Array Detector and compared with the standard values obtained from the Radiation Field Analyzer (RFA) during Commissioning

Read more

Summary

Introduction

Flatness and symmetry of a radiation beam: The flatness of the beam is defined by the following formula: Flatness (%) = Dmax/Dmin × 100% (1). Radiation field symmetry is defined as the maximum ratio of doses at two symmetric points relative to the central axis of the field. In the treatment of superficial cancers electron beam therapy is globally used. Due to the unique characteristics of an electron beam it is suitable for the treatment of disease near to the skin surface and sparing the deep tissues. To conform the shape of the radiation field to that of the irregular shapes of individual tumours, and sparing peripheral tissues ,the need for electron cut-outs is required [1]. The goal of this study is to compare the electron beam characteristics, such as Flatness and Symmetry. The flatness of the beams may change significantly with depth since scattered electrons are of lower energy

Methods
Results
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.