Abstract

Detection of Neisseria gonorrhoeae relies increasingly on nucleic acid amplification tests (NAATs). The specificity of many gonococcal NAATs has been suboptimal and supplementary testing remains recommended in Europe and several additional countries. The novel dual-target GeneProof Neisseria gonorrhoeae PCR kit, targeting porA pseudogene and 16S rRNA gene, showed a high specificity and sensitivity when isolates of non-gonococcal Neisseria and related species (n = 144), and gonococci (n = 104) were tested. However, rare gonococcal porA mutants were only detected in the 16S rRNA gene target and two non-gonococcal isolates showed a low-level cross-reactivity in the 16S rRNA gene target. The detection limit for both targets was 1.5 copies per reaction.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.