Abstract

In this contribution, the analytical potential of total reflection X-ray fluorescence (TXRF) instrumentation has been evaluated for the determination of major and trace elements in milk powder. TXRF allows the possibility of direct analysis of solid suspensions without the need for a digestion process and therefore it can be a potential analytical candidate for simple and cost-effective analysis.A detailed study to select sample preparation and measurements conditions was carried out. Different quantification approaches (including internal standardization and empirical calibration) were also tested. Finally, the developed TXRF methods (W anode) were validated by a strict comparison with the data from the reference methods on a set of twenty-three samples using robust statistics. Results showed that acceptable results can be obtained for K, Ca, Fe and Zn determination if using adequate calibration approaches. Otherwise, only screening results can be obtained for light elements (P and Cl) in milk powder samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.