Abstract

Recent progress in the development of ion trap ICP-MS instrumentation is presented in terms of performance figures of merit for trace elemental and isotopic analysis. Conventional polyatomic ion interferences are avoided altogether using collisional dissociation techniques inherent to ion trap operation. The current detection limits are in the low pg mL –1 range, as obtained over extended mass scans using multi-element solutions. The dynamic range linearity is nearly six orders of magnitude using variable injection times to control ion density in the trap. The abundance sensitivity is 10 5 to 10 6 using selective ion monitoring. Other performance figures approach or exceed those of conventional linear quadrupole based ICP-MS instrumentation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.