Abstract
The portable blood gas analyzer OPTI Critical Care Analyzer was evaluated in comparison to routine laboratory assays using heparinized blood samples of adults and newborns. Within-run imprecision studies were performed with native blood using tonometry to adjust blood gas concentrations. The results obtained show a very close agreement between the OPTI system and the comparison methods for all parameters tested: hemoglobin (y=1.00x−0.2 g/l, r=0.997, n=81), sodium (y=1.13x−18.7 mmol/l, r=0.951, n=79), potassium (y=1.03x−0.04 mmol/l, r=0.972, n=79), pH (y=1.03x−0.29, r=0.958, n=57), pCO2 (y=1.03x−1.14 mm Hg, r=0.988, n=57) and pO2 (y=1.07x−0.85 mm Hg, r=0.995, n=57). The coefficients of variation for the within-run imprecision were below 1.1% for sodium and hemoglobin, and below 2.6% for all other parameters, except for pCO2 with coefficients of variation up to 3.6% at low calibration gas concentrations. Due to this analytical performance and its portability, the OPTI system is well suited for low to medium test frequencies and immediate use in emergency rooms, intensive care or surgery units.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.