Abstract

A new analytical technique has been developed that measures both total protium (H) and deuterium (D) in a gas mixture containing H2, D2, and HD. This new analytical technique uses a micro gas chromatograph (GC) with two molecular sieve columns. One column uses D2 as the carrier gas and the other uses H2 as the carrier gas. Laboratory tests have shown that when used in this configuration the GC can measure both total protium and total deuterium, each with a sensitivity of less than 20 ppm. This new analytical technique was developed as a result of a request to provide instrumentation to measure the protium and deuterium concentrations at several process points during initial testing of the new hydrogen tritium thermal cycling absorption process columns.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.