Abstract
To estimate Teneligliptin by Ultra-Fast Liquid Chromatographic (RP-UFLC), a fast, reliable, and cost-effective UFLC method was developed. The approximate flowing rate was 1.0 mL/minfor the Phenomenex Kinetex column C18 (250mmx 4.6mm. 5 Micron) with MeOH and ACN (60:40 v/v) as a portable stream. A PDA detector at 246 nm was used to track the eluent. Teneligliptin elutes at 4.938 minutes of the processing period. The proposed method has a linearity of 10 to 50μg/ml concentration range; the calibration diagram is continuous and 0.9942 percent for the correlations variable (r2). For the suggested procedure, both LOD and LOQ are measured as 0.097 and 1.023μg/mL, respectively. In the pharmaceutical product, the drug has been under the states of stress conditions acidic, alkaline, degradation, Ultraviolet, and heat stress. The degradation product from the Teneligliptin was well solved, and thus the stability of this method was demonstrated. The process was tested in compliance with ICH criteria of device suitability, linearity, reliability, consistency, and robustness.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.