Abstract

Objective of the present work is to develop and validate a simple, cost effective, sensitive and fast HPLC method for the analysis of Secnidazole. A Shimadzu HPLC system with Luna 5µm C18 column is employed for the analysis using Methanol:H2O (60:40, v/v) as mobile phase. Signal from Secnidazole is detected at 310nm by UV Spectrophotometer. The proposed method is fully validated and found to be linear over a workable drug concentration, accurate, precise and robust. This fast and inexpensive method is suitable for research laboratories as well as for quality control analysis in pharmaceutical industries.Sharmin et al., International Current Pharmaceutical Journal, March 2016, 5(4): 41-44http://www.icpjonline.com/documents/Vol5Issue4/03.pdf

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.