Abstract
A dry film resist (DFR) chip compatible with the Agilent Bioanalyzer 2100 was designed and fabricated for use in the analysis of lactate in serum by chip isotachophoresis (ITP). The Agilent Bioanalyzer 2100 is a commercially available field deployable analytical instrument originally developed for the electrophoretic analysis of DNA, RNA and proteins. The DFR chip was designed for the ITP separation of lactate in human serum within 1min and was made compatible with the Bioanalyzer after packaging in the plastic caddies normally used for the DNA chips. A 20-fold improvement in sensitivity was obtained for the DFR chips in comparison with the standard chips used in earlier work. The limit of detection and limit of quantification for lactate were 24μM and 80μM, respectively. This new approach enables the use of commercial platforms like the Agilent Bioanalyzer for new applications including the analysis of small molecules.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.