Abstract

To determine the analytical performance of Novel VITROS BRAHMS Procalcitonin Immunoassay on VITROS 3600 and correlation with BRAHMS PCT sensitive KRYPTOR reference method. Analytical performances including imprecision studies, linearity, limit of detection (LoD) and determination of hemolysis index were performed for VITROS BRAHMS PCT assay. Imprecision was assessed on plasma pool and internal control with 2 levels. The method comparison was performed using 162 plasma obtained from clinical departments. The total imprecision was acceptable and all CV were <5%. The LoD was in accordance with manufacturer’s claims. The equation of linearity in the lower range was found to be y = 1.0014x – 0.0091, with r2 = 1. No interference to hemoglobin up to 11 g/L was observed. Correlation studies showed a good correlation between PCT measurements using VITROS BRAHMS PCT assay against KRYPTOR system including for values lower than 2 µg/L. The novel VITROS BRAHMS PCT assay from OrthoClinical Diagnostics shows analytical performances acceptable for clinical use. In addition, the concordance with KRYPTOR method was fine at all clinical cut-offs.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.