Abstract

In radiation therapy, estimation of surface doses is clinically important. This study aimed to obtain an analytical relationship to determine the skin surface dose, kerma and the depth of maximum dose, with energies of 6 and 18 megavoltage (MV). To obtain the dose on the surface of skin, using the relationship between dose and kerma and solving differential equations governing the two quantities, a general relationship of dose changes relative to the depth was obtained. By dosimetry all the standard square fields of 5x5cm to 40x40cm, an equation similar to response to differential equations of the dose and kerma were fitted on the measurements for any field size and energy. Applying two conditions: a) equality of the area under dose distribution and kerma changes in versus depth in 6 and 18 MV, b) equality of the kerma and dose at x=dmax and using these results, coefficients of the obtained analytical relationship were determined. By putting the depth of zero in the relation, amount of PDD and kerma on the surface of the skin, could be obtained. Using the MATLAB software, an exponential binomial function with R-Square >0.9953 was determined for any field size and depth in two energy modes 6 and 18MV, the surface PDD and kerma was obtained and both of them increase due to the increase of the field, but they reduce due to increased energy and from the obtained relation, depth of maximum dose can be determined. Using this analytical formula, one can find the skin surface dose, kerma and thickness of the buildup region.

Highlights

  • By dosimetry all the standard square fields of 5cm×5cm to 40cm×40cm, an equation similar to response to differential equations of the dose and kerma were fitted on the measurements for any field size and energy

  • Using the MATLAB software, an exponential binomial function with R-Square >0.9953 was determined for any field size and depth in two energy modes 6 and 18MV, the surface percentage depth dose (PDD) and kerma was obtained and both of them increase due to the increase of the field, but they reduce due to increased energy and from the obtained relation, depth of maximum dose can be determined

  • High-energy photons generated by a medical linear accelerators are used in radiation therapy

Read more

Summary

Introduction

High-energy photons generated by a medical linear accelerators are used in radiation therapy. With an increasing energy of these radiations, the penetration power of photons and secondary electrons will be increased and as a result the point of maximum dose (d_max) is placed in more depth (Apipunyasopon et al, 2013).Depending on the amount of energy, dose build up region occurs for x-rays megavoltage in the first depths of the entrance surface of phantom or skin. This study aimed to obtain an analytical relationship to determine the skin surface dose, kerma and the depth of maximum dose, with energies of 6 and 18 megavoltage (MV). Results: Using the MATLAB software, an exponential binomial function with R-Square >0.9953 was determined for any field size and depth in two energy modes 6 and 18MV, the surface PDD and kerma was obtained and both of them increase due to the increase of the field, but they reduce due to increased energy and from the obtained relation, depth of maximum dose can be determined. Conclusions: Using this analytical formula, one can find the skin surface dose, kerma and thickness of the buildup region

Objectives
Methods
Results
Conclusion

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.