Abstract

A method for multielemental (Ca, Cr, Cu, Fe, K, Mg, Mn, Na, P and Zn) determination in multimineral/multivitamins by atomic emission spectrometry in a medium power radiofrequency capacitively coupled plasma (275 W) and low Ar consumption (0.4 L min(-1)) is proposed. Determinations were performed on commercially available tablets and a standard reference material after acidic high-pressure microwave assisted digestion and using the standard additions procedure. The detection limits (mg g(-1)) were in the range 0.003 (Na)-1.5 (P) and were not depreciated by the non-spectral interference of mineral matrices of K, Ca, Mg and Na excepting Zn and P. Found concentrations corresponded generally to the labelled contents with recovery in the range of 90-107% and 1.0-13.0% repeatability. The proposed technique could be an advantageous alternative to the more expensive inductively coupled plasma atomic emission spectrometry in the quality control of multimineral/multivitamin preparations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.