Abstract
Knowledge of the geometrical solid angle is essential in all absolute measurements of the strengths of radioactive materials and to calibrate detectors. A direct mathematical formalism for the determination of the geometrical solid angle and the geometrical efficiency of a circular detector and an arbitrarily positioned line source is deduced. The results have been compared with previous computational treatments. The comparison shows a very satisfactory agreement in all cases.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Scientific Publications of the State University of Novi Pazar Series A: Applied Mathematics, Informatics and mechanics
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.