Abstract
Suppression or enhancement of MS ionization, particularly evident when electrospray is used as the source of ions, has been widely discussed. An assay for a small-molecule pharmaceutical in dog plasma between 1-300 ng/ml was validated with a mean bias across the calibration range of 5.0%. When the calibration sample matrix was substituted for human plasma, the mean bias across the range increased to 29.1%. A study of bias originating as a result of matrix effects, arising from endogenous glycerophosphocholine species, in plasma sources is discussed. A simple strategy to assess the potential of any unmitigated matrix effect to bias quantitative analysis by nonequivalent ionization induction or suppression is evaluated.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.