Abstract

The HPGe detector efficiency is measured as a function of source to detector separation using disc sources of 131I with diameter ranging from 10 to 400mm. Detector efficiencies are characterized using single photon point-like standard sources at different distances; the calculated efficiencies for disc sources were analyzed by utilizing the double point detector model (DPDM) and the efficiency transfer method. The developed approach provided satisfactory results. The axial variation and radial dependence for disc sources efficiency determination in gamma-ray spectrometry were described with both gamma ray standard sources and measured samples as their extended sources.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.