Abstract

In this study, detection sensitivity of the conventional flame atomic absorption spectrophotometer (FAAS) for the determination of manganese (Mn2+) was enhanced by employing a preconcentration method from wastewater samples. Flower-shaped Ni(OH)2 nanomaterials were synthesized and used as sorbent material in preconcentration procedure. With the aim of attaining optimum experimental conditions, effective parameters of extraction method were optimized and these included pH of buffer solution, desorption solvent concentration and volume, mixing type and period, nanoflower amount, and sample volume. The detection limit of the optimized method was determined to be 2.2 μg L-1, and this correlated to about 41-fold enhancement in detection power relative to direct FAAS measurement. Domestic wastewater was used to test the feasibility of the proposed method to real samples by performing spike recovery experiments. The wastewater sample was spiked at four different concentrations of manganese, and the percent recoveries determined were in the range of 95-120%.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call