Abstract
LDL-cholesterol (LDL-C) concentration is currently determined in most clinical laboratories by the Friedewald calculation. This approach has several limitations and may not meet the current total error requirement in LDL-C measurement of < or = 12%. We evaluated the analytical and clinical performance of the direct N-geneous LDL-C assay (Equal Diagnostics). The N-geneous method correlated highly with the modified beta-quantification assay (r = 0.95; y = 0.91x + 70.6 mg/L; n = 199), showed no significant effect of increased triglyceride or other common interferants, and performed adequately in serum samples from nonfasting individuals. This assay demonstrated a mean total error of 6.75% over a wide range of LDL-C concentrations. In addition, at the medical decision cutoff points, this LDL-C assay showed positive predictive values of 78-95% and negative predictive values of 84-99%. We conclude that the N-geneous LDL-C meets the currently established analytical performance goals and appears to have a role in the diagnosis and management of hypercholesterolemic patients.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.