Abstract

The quantification of target analytes in complex matrices requires special calibration approaches to compensate for additional capacity or activity in the matrix samples. The standard addition is one of the most important calibration procedures for quantification of analytes in such matrices. However, this technique requires a great number of reagents and material, and it consumes a considerable amount of time throughout the analysis. In this work, a new calibration procedure to analyze biological samples is proposed. The proposed calibration, called the addition calibration technique, was used for the determination of zinc (Zn) in blood serum and erythrocyte samples. The results obtained were compared with those obtained using conventional calibration techniques (standard addition and standard calibration). The proposed addition calibration was validated by recovery tests using blood samples spiked with Zn. The range of recovery for blood serum and erythrocyte samples were 90-132% and 76-112%, respectively. Statistical studies among results obtained by the addition technique and conventional techniques, using a paired two-tailed Student's t-test and linear regression, demonstrated good agreement among them.