Abstract

The Georgia Bureau of Investigation utilized a silylation method of analysis for low molecular weight carboxylic acids in the past. Due to the negative impact such derivatizations can have on gas chromatography-mass spectrometry (GC-MS) systems an alternative means of analysis was investigated. The described method is a whole blood solid phase extraction of valproic acid, salicylic acid and ibuprofen utilizing butylation for sensitivity and improved chromatography by GC-MS. The method produced a limit of detection and limit of quantitation at 1 mg/L for valproic acid, 2 mg/L for salicylic acid and 0.25 mg/L for ibuprofen. The variability based upon the middle of the calibration curve estimated to be 7% for valproic acid, 8% for salicylic acid and 11% for ibuprofen established upon a 95% confidence interval, with the highest percent coefficient of variation being 5.3% for ibuprofen.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.