Abstract

In order to optimise the analysis of urinary nucleosides by high performance liquid chromatography/mass spectrometry (HPLC/MS), the HPLC separation of these compounds was performed at different 'flow rates' and 0.2mL/min was found to give both a better separation and ionisation. The ionisation conditions were optimised to give the best intensity of the molecules quasi-molecular ions. The ion distribution profile and ionisation in both positive and negative mode were examined and the detection of the protonated molecule in positive mode chosen for further analysis. The limits of detection of the method developed are reported and representative LC/MS and LC/MS/MS spectra shown. Typical urinary nucleoside chromatograms are presented.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.