Abstract

To further characterize the kinking forces and degree of curvature that result in failure of various stents, as malignant obstruction of the ureter causes failure of about half of ureteric stents. Several stents (Silhouette 4.6 F and 6 F, Applied Medical, Rancho Santa Margarita, CA; Sof-Curl Tecoflex 6 F, Gyrus ACMI, Southborough, MA; Polaris Ultra 6 and 7 F, and Percuflex 6, 7 and 8 F stents, Boston Scientific, Natick, MA, USA) were tested. The amount of force necessary to result in kinking of the stent was measured, and the degree of curvature at failure was calculated for each stent. The Silhouette 4.6 and 6 F stents were the most resistant to failure by kinking and curvature. In general, smaller stents allowed more curvature before failing than their larger counterparts. The greater allowable curvature and resistance to kinking achieved by the Silhouette ureteric stents might result in fewer stent failures in cases of malignant obstruction or other retroperitoneal processes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.