Abstract
A series of experiments was performed to analyze the utility of capillary electrophoresis (CE) with multiwavelength detection capabilities for multiplex typing of short tandem repeat loci. Characteristics of the sieving polymer, hydroxyethylcellulose, which affect resolution of single strand (ss) DNA fragments were examined. Additionally, the effects of denaturant in the polymer system, separation voltage, and analysis temperature were studied to ascertain their effects on DNA separations and capillary lifetime. The use of elevated run temperature (60 degrees C) was found to improve sizing precision, to increase the lifetime of capillaries (100 runs or more per capillary), and to provide runtimes of under 20 min. Finally, 100 individual human DNA samples were typed successfully using CE. The average resolution obtained was 1.4 bases for a 200 base fragment with a standard deviation of sizing of 0.2 bases, allowing all alleles examined to be distinguished clearly.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.