Abstract
The purpose of this study was to analyse the anatomy of the trapezium with regard to cup position in trapeziometacarpal replacement and identify those trapeziums that are at risk of cup perforation through the trapezoid articular surface. The width of the proximal and distal articular surface and the height of the trapezium and second metacarpal facet were measured on 96 peritrapezial views reconstructed from computed tomography scans. The trapezoid articular surface of the trapezium (TRAST) angle was calculated, and four different cup designs were virtually positioned centrally in the trapezium and parallel to the proximal articular surface. Risk of perforation was defined as a cup that exceeds the ulnar border of the proximal articular surface. The mean TRAST angle in our study was 33°. Risk of perforation is higher when the second metacarpal facet is smaller than 5 mm and when the TRAST angle is 35° or more. In these cases, the position of the cup should be more radial or more distal.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.