Abstract

The goal of this study is to compare the calibration factors obtained by the four calibration methods recommended by the standard ISO 8529-2. To do this, two Berthold LB6411 neutron monitors are calibrated applying the four methods using a 241Am-Be source and the calibration results are compared. Calibration factors obtained are coherent with each other considering their typical standard deviation and differ by less than 4% from each other.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.