Abstract

Based on the principle of the Michelson interferometer, the paper briefly describes the theoretical significance and calculates and deduces three expressions of the interference modulation depth. The influence of the surface shape error of plane mirror on modulation depth is analyzed, and the tolerance of error is also pointed out. Moreover, the dependence of modulation depth on the reflectance change of beam splitter interface is also analyzed, and the curve is given. It is concluded that this paper is of general significance for the Fourier transform spectrometer based on the principle of the Michelson two-beam interference.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.