Abstract

In the present research, the performance of three sample injection devices in a portable capillary electrophoresis (CE) instrument was examined. These were the so-called cross-sampler, horizontal injection channel and vertical injection channel. All the three showed a good reproducibility of migration times (the relative standard deviation (RSD) was 4.3% in the case of the cross-sampler, 6.0% in the case of the horizontal injection channel and 1.7% in the case of the vertical channel). However, the reproducibility of peak areas was not sufficient. Hence, this study was mainly focused on qualitative analysis. The cross-sampler injection device was used in the portable CE instrument to analyse the composition of degradation products of chemical warfare agents (CWA). For the analysis of CWA degradation products simple procedures for the extraction of phosphonic acids from different surfaces, such as soil, concrete and granite blocks, tile floor, were developed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.