Abstract

Analysis procedures and instrumentation to measure low levels of sulfur hexafluoride in seawater are described. The minimum detectable level for a 500 mL sample is about 3×10−17 mol/L. Concentrations from 1.5 to 300×10−15 mol/L can be measured with approximately 2% precision. The procedure includes a concentration step employing a trap of Porapak‐Q© in a dry ice/2‐propanol bath. The system has been used for tracer experiments in Santa Monica Basin and Santa Cruz Basin, and will be improved for future tracer release experiments in the deep ocean.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.