Abstract

A recently described thermospray flame furnace atomic absorption spectrometric (TS-FF-AAS) system has been modified in order to extend the applicability of the method for the determination of elemental traces in very small sample volumes (microliter or submicroliter). As an easily available, effective thermospray vaporizer, a fused silica capillary was used and the liquid sample was transported by 1 MPa (10 bar) gas pressure delivered by a standard gas cylinder. A 0.3 microL sample volume can be analyzed with a higher power of detection than using 3 orders of magnitude larger sample volumes with conventional flame atomic absorption spectrometry. The relative standard deviations (N=12) for 0.3 microL volumes and 5 microg/mL Pb samples amount to 3.1% and 3.8% in signal height and signal area, respectively. The detection limit was found to be 69 ng/mL. Initial experiments with other elements (Cd, Hg, Tl, Zn) led to similar results.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.