Abstract

A multiple ion-counting total evaporation technique has been developed for precise analysis of sub-picogram (pg) quantities of 238Pu by thermal ionization mass spectrometry (TIMS). A resin bead loading procedure was tailored to minimize uranium (U) contamination during loading. This loading technique, coupled with a mass spectrometry protocol that minimizes U ionization and removal of bulk U with ion exchange chromatography, allows for single pg-sized aliquots of plutonium (Pu) to be measured without interference from 238U. This technique has the potential to provide complete isotopic characterization of trace amounts of Pu by TIMS, without the need for alpha spectrometry.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.