Stray field outside of magnetic resonance imaging (MRI) magnet must be confined to five gauss in a limited region. It's hard to high field MRI system. In the paper, calculation of magnetic field of MRI magnet is studied Approximate expression of stray field outside of MRI magnet is deduced Influence of shielding coil position on stray field is analyzed and an expression determining shielding coil position is given to cancel stray field of negative five orders.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.