Abstract

Although graphite furnace atomic absorption spectrometry is one of the most sensitive techniques for trace element determinations, the analysis of solid materials can be challengingly difficult. As no dissolution is used, none of the sample matrix is removed before the introduction of the sample into the furnace, and this can result in severe vapour-phase interferences. Typically, quantification has been achieved by the use of standard additions or by comparison with a known reference material. Using stabilised temperature platform furnace technology and simple aqueous standards, we have determined Cr in plastic film, Pb in flexible PVC and Cu in NBS Standard Reference Material bovine liver. Agreement was obtained with the values provided by the sample suppliers. The precision was found to be about 4–6%(relative standard deviation).The solid materials analysed contained relatively high levels of analyte, requiring the use of alternative wavelengths and in two instances an increased gas flow during atomisation. The effect of gas flow on sensitivity was examined and 19 wavelengths were characterised for use with Zeeman background correction.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.