Abstract

As a part of QA/QC of k0-based internal monostandard neutron activation analysis (IM-NAA), three types of synthetic multielement standards (SMELS) were analyzed using pneumatic fast transfer system irradiation position of KAMINI reactor, IGCAR. Radioactive assay of activation products was carried out by high resolution gamma ray spectrometry. IM-NAA was used to determine relative concentration ratios of 22 elements with respect to gold internal monostandard. Absolute concentrations were calculated using assigned concentration of Au in all the types of SMELS. Z-score values within ±1 at 95.5 % confidence level and percentage deviations within ±5 % indicated good quality of the results by IM-NAA in most of the cases. Using this methodology, an ilmenite mineral sample was analyzed and concentrations of 14 elements were determined using Sc as monostandard.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.