Abstract

The feasibility of ultra thin-layer chromatography atmospheric pressure matrix-assisted laser desorption ionization mass spectrometry (UTLC-AP-MALDI-MS) has been studied in the analysis of small molecules. Because of a thinner adsorbent layer, the monolithic UTLC plates provide 10-100 times better sensitivity in MALDI analysis than conventional high performance thin-layer chromatography (HPTLC) plates. The limits of detection down to a low picomole range are demonstrated by UTLC-AP-MALDI-MS. Other advantages of UTLC over HPTLC include faster separations and lower solvent consumption. The performances of AP-MALDI-MS and vacuum MALDI-MS have been compared in the analysis of small drug molecules directly from the UTLC plates. The desorption from the irregular surface of UTLC plates with an external AP-MALDI ion source combined with an ion trap instrument provides clearly less variation in measurements of m/z values when compared with a vacuum MALDI-time-of-flight (TOF) instrument. The performance of the UTLC-AP-MALDI-MS method has been applied successfully to the purity analysis of synthesis products produced by solid-phase parallel synthesis method.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call