Micro-determination of serotonin and histamine was accomplished by means of high-speed liquid chromatography and automated fluorometric analysis using 0- phthaldialdehyde. 0.5 ng of serotonin and 0.1 ng of histamine can be easily deter-mined by the methods described here. For the analysis of urine of healthy human subject, any pre-purification was not carried out to measure serotonin, but in the histamine analysis, extraction by a column of weak cation exchange resin was necessary before injection to the high-speed liquid chromatograph. It was found that diurnal rhythm of urinary serotonin was statistically significant (p<0.01) in healthy human subject, and that the night level of serotonin was slightly higher than the daytime level.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.