Abstract

Risperidone is currently one of the most frequently prescribed atypical antipsychotic drugs; its main active metabolite 9-hydroxyrisperidone contributes significantly to the therapeutic effects observed. An original analytical method is presented for the simultaneous analysis of risperidone and the metabolite in plasma, urine and saliva by high-performance liquid chromatography coupled to an original sample pre-treatment procedure based on micro-extraction by packed sorbent (MEPS). The assays were carried out using a C8 reversed-phase column and a mobile phase composed of 73% (v/v) acidic phosphate buffer (30mM, pH 3.0) containing 0.23% triethylamine and 27% (v/v) acetonitrile. The UV detector was set at 238nm and diphenhydramine was used as the internal standard. The sample pre-treatment by MEPS was carried out on a C8 sorbent. The extraction yields values were higher than 92% for risperidone and 90% for 9-hydroxyrisperidone, with RSD for precision always lower than 7.9% for both analytes. Limit of quantification values in the different matrices were 4ng/mL or lower for risperidone and 6ng/mL or lower for the metabolite. The method was successfully applied to plasma, urine and saliva samples from psychotic patients undergoing therapy with risperidone, with satisfactory accuracy results (recovery>89%) and no interference from other drugs. Thus, the method seems to be suitable for the therapeutic drug monitoring of schizophrenic patients using the three different biological matrices plasma, urine and saliva.