Abstract

Proportional counters and a β-γ coincidence system and their performance characteristics are described, with emphasis on internal and external quality controls. Typical limits of detection (LD) are 8mBqm−3 for 85Kr and 4mBqm−3 for 133Xe for the proportional counters, while for similar xenon volumes the LD for 133Xe in the β-γ system is approximately 4 times lower. Results of comparative analyses agree above the limit of quantification. Xenon-133 levels in southern Germany are presently at or below the LD of the proportional counters, but above the LD of the β-γ system, which is now routinely used for analysis of environmental samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.