Abstract

The Monte Carlo method is widely used in the Gamma Knife dose distribution calculations. In this study, Monte-Carlo simulation with Geant4 was applied to determine Leksell Gamma Knife dose distribution for homogeneous solid water and heterogeneous brain phantoms, and primarily focused on the differences arising from the effects of inhomogeneity. The relative dose graphs were displayed for different collimator diameters. The dose values from the maximal interval regions were determined and compared. It seen that when the collimator size was 4mm, the mean maximum relative dose point was at 47mm for water phantom and at 44mm for the brain. The difference found was about 6.38%. The differences among the results determined for water and brain were ranged from 2 to 17%. As a result, the inhomogeneity effect should be considered in dose calculation in Gamma Knife planning system.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.